Skip to main content
Orthopedics, 2012;35(4):e561–e565
Published Online:https://doi.org/10.3928/01477447-20120327-29Cited by:13

Abstract

This article describes a case of late locking mechanism failure and disassociation of a Pinnacle acetabular cup (DePuy, Warsaw, Indiana) and Marathon polyethylene liner (DePuy) 53 months after routine primary total hip arthroplasty in an active patient. Following an uncomplicated initial postoperative recovery, the patient felt a pop while kneeling for gardening activities. The patient had no prodromal symptoms but reported pain with range of motion and weight bearing following the episode. Radiographs appeared to show a fractured liner with intact acetabular and femoral components. Intraoperatively, the liner was disassociated and dislocated inferior to the acetabulum, and 3 consecutive antirotational tines were sheared off the liner at their bases. The acetabular cup and femoral stem were well fixed. Extensive metallosis existed from the femoral head and acetabular shell articulating in the absence of a liner. The patient underwent revision of the acetabular cup, polyethylene liner, and femoral head due to the concern for osteolysis given the extensive metallosis. Revision of the acetabular cup and liner were performed with no further complications. No conclusions could be made as to the exact mechanism of failure following laboratory analysis of the retrieved polyethylene liner. Several possible mechanisms of failure are possible. Although cases of acute disassociation of this system have been reported, this is the first article to our knowledge to report failure at such a late postoperative time.

  • 1.Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999; 46(4):573–581.10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A

    > Crossref MedlineGoogle Scholar
  • 2.Jacobs CA, Christensen CP, Greenwald AS, McKellop H. Clinical performance of highly cross-linked polyethylenesin total hip arthroplasty. J Bone Joint Surg Am. 2007; 89(12):2779–2786.10.2106/JBJS.G.00043

    > Crossref MedlineGoogle Scholar
  • 3.Bradford L, Baker D, Ries MD, Pruitt LA. Fatigue crack propegation resistance of highly crosslinked polyethylene. Clin Orthop Relat Res. 2004; (429):68–72.10.1097/01.blo.0000150124.34906.34

    > Crossref MedlineGoogle Scholar
  • 4.Ries MD. Review of the evolution of the cementless acetabular cup. Orthopedics. 2008; 31(12 suppl 2):88–91.

    > MedlineGoogle Scholar
  • 5.Mesko JW. Acute liner dissassociation of a Pinnacle acetabular component [published online ahead of print June 13, 2008]. J Arthroplasty. 2009; 24(5):815–818.10.1016/j.arth.2008.03.010

    > Crossref MedlineGoogle Scholar
  • 6.Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty. 2009; 24(1):158.e15–158.e19.10.1016/j.arth.2007.12.020

    > CrossrefGoogle Scholar
  • 7.Powers CC, Fricka KB, Austin MS, Engh CA. Five Duraloc locking ring failures [published online ahead of print January 22, 2010]. J Arthroplasty. 2010; 25(7):1170.e15–1170. e18.10.1016/j.arth.2009.10.013

    > CrossrefGoogle Scholar
  • 8.Werle J, Goodman S, Schurman D, Lannin J. Polyethylene liner dissociation in Harris-Galante acetabular components: a report of 7 cases. J Arthroplasty. 2002; 17(1):78–81.10.1054/arth.2002.29327

    > Crossref MedlineGoogle Scholar
  • 9.Saito S, Ryu J, Seki M, Ishii T, Saigo K, Mori S. Analysis and results of dissociation of the polyethylene liner in the Harris-Galante I acetabular component [published online ahead of print October 23, 2007]. J Arthroplasty. 2008; 23(4):522–526.10.1016/j.arth.2007.06.004

    > Crossref MedlineGoogle Scholar
  • 10.Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996; (333):87–95.

    > MedlineGoogle Scholar
  • 11.Rosca M, Postak PD, Greenwald AS. Enhanced cross-linked polyethylene modular acetabular designs: performance characteristics. Orthopaedic Reserach Laboratories Web site. http://orl-inc.com/hip_publications/2008/crosslinkedPolyCup.pdf. Updated 2008. Accessed March 26, 2011.

    > Google Scholar
  • 12.Tradonsky A, Postak SD, Froimson AI, Greenwald AS. A comparison of the disassociation strength of modular acetabular components. Clin Orthop Relat Res. 1993; (296):154–160.

    > MedlineGoogle Scholar
  • 13.Engh CA, Hopper RH, Engh CA. Long-term porous-coated cup survivorship using spikes, screws, and press-fitting for initial fixation. J Arthroplasty. 2004; 19(7 suppl 2):54–60.10.1016/j.arth.2004.06.004

    > Crossref MedlineGoogle Scholar
  • 14.Udomkiat P, Dorr LD, Wan Z. Cementless hemisphereic porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am. 2002; 84(7):1195–2000.

    > Crossref MedlineGoogle Scholar
  • 15.Ries MD. Review of the evolution of the cementless acetabular cup. Orthopedics. 2008; 31(12 suppl 2):88–91.

    > MedlineGoogle Scholar
  • 16.Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007; 89(10):2212–2217.10.2106/JBJS.F.00758

    > Crossref MedlineGoogle Scholar
  • 17.Furmanski J, Atwood SA, Bal BS, et al.Fracture of Highly Crosslinked UHMWPE Acetabular Liners. Orthopaedic Reserach Laboratories Web site. http://orl-inc.com/hip_publications/2008/LinerFractureJevan.pdf. Updated 2008. Accessed March 26, 2011.

    > Google Scholar
  • 18.Callaghan JJ, Rosenberg AG, Rubash HE. The Adult Hip. 2nd ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2007.

    > Google Scholar
  • 19.Kindsfater K, Barrett W, Dowd J, South-worth C, Cassell M. Midterm survival of the Pinnacle multi-liner acetabular cup in a prospective multi-center study. Poster presented at: American Academy of Orthopaedic Surgeons Proceedings; February 16, 2007; San Diego, California. Poster No. PO77.

    > Google Scholar
  • 20.Powers CC, Ho H, Beykirch SE, et al.A Comparison of a second- and a third-generation modular cup design: is new improved [published online ahead of print April 9, 2009]?J Arthroplasty. 2010; 25(4):514–521.10.1016/j.arth.2009.02.018

    > Crossref MedlineGoogle Scholar

We use cookies on this site to enhance your user experience. For a complete overview of all the cookies used, please see our privacy policy.

×