Abstract
PURPOSE:
To compare vault height predictability of Implantable Collamer Lens (ICL; Staar Surgical) sizing using a sulcus diameter–based formula or the manufacturer-recommended white-to-white–based method.
METHODS:
In 50 myopic eyes, ICL size was calculated using both a formula including sulcus diameter and the traditional formula based on white-to-white diameter. Sulcus diameter was measured using Artemis 2 very high-frequency (VHF) digital ultrasound (ArcScan Inc). Implantation was based on the sulcus diameter derived size. Actual postoperative vault height achieved was measured by VHF digital ultrasound scanning. Circle segment trigonometry was used to calculate the vault height that would have resulted had lens sizing been based on the white-to-white formula.
RESULTS:
The same lens size would have been used in 60% of eyes, a smaller lens would have been used in 34% of eyes and a larger lens in 6% of eyes had lens sizing been based on the white-to-white formula. Mean vault for eyes with lenses sized using the sulcus diameter formula was 0.37±0.16 mm (range: 0.08 to 0.92 mm), with 2% <0.09 mm, the recognized low-vault height for risk of cataract. Circle segment trigonometry predicted that the vault height would have been 0.24±0.28 mm (range: −0.31 to 0.92 mm), with 26% <0.09 mm had lens sizing been based on the white-to-white formula.
CONCLUSIONS:
Significantly better predictability of postoperative vault height was achieved by including sulcus diameter into the ICL sizing formula compared with using the traditional white-to-white–based formula.
- 1.Sanders DR, Vukich JA. Incidence of lens opacities and clinically significant cataracts with the implantable contact lens: comparison of two lens designs. J Refract Surg. 2002; 18(6):673–682. Erratum in J Refract Surg. 2003;19(4):2nd page following table of contents. LinkGoogle Scholar
- 2.Sanders DR, Vukich JA, Doney K, Gaston M. U.S. Food and Drug Administration clinical trial of the Implantable Contact Lens for moderate to high myopia. Ophthalmology. 2003; 110(2):255–266.
10.1016/S0161-6420(02)01771-2 Crossref MedlineGoogle Scholar - 3.Kodjikian L, Gain P, Donate D, Rouberol F, Burillon C. Malignant glaucoma induced by a phakic posterior chamber intraocular lens for myopia. J Cataract Refract Surg. 2002; 28(12):2217–2221.
10.1016/S0886-3350(02)01213-0 Crossref MedlineGoogle Scholar - 4.Reed JE, Thomas JV, Lytle RA, Simmons RJ. Malignant glaucoma induced by an intraocular lens. Ophthalmic Surg. 1990; 21(3):177–180. LinkGoogle Scholar
- 5.Fernandes P, Gonzalez-Meijome JM, Madrid-Costa D, Ferrer-Blasco T, Jorge J, Montes-Mico R. Implantable collamer posterior chamber intraocular lenses: a review of potential complications. J Refract Surg. 2011; 27(10):765–776.
10.3928/1081597X-20110617-01 LinkGoogle Scholar - 6.Brandt JD, Mockovak ME, Chayet A. Pigmentary dispersion syndrome induced by a posterior chamber phakic refractive lens. Am J Ophthalmol. 2001; 131(2):260–263.
10.1016/S0002-9394(00)00606-1 Crossref MedlineGoogle Scholar - 7.Kohnen T, Kasper T, Terzi E. Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation [German]. Ophthalmologe. 2005; 102:1105–1117.
10.1007/s00347-005-1274-7 Crossref MedlineGoogle Scholar - 8.Mastropasqua L, Toto L, Nubile M, Falconio G, Ciancaglini M. Long-term complications of bilateral posterior chamber phakic intraocular lens implantation. J Cataract Refract Surg. 2004; 30(4):901–904.
10.1016/j.jcrs.2003.08.012 Crossref MedlineGoogle Scholar - 9.Brauweiler PH, Wehler T, Busin M. High incidence of cataract formation after implantation of a silicone posterior chamber lens in phakic, highly myopic eyes. Ophthalmology. 1999; 106(9):1651–1655.
10.1016/S0161-6420(99)90352-4 Crossref MedlineGoogle Scholar - 10.El-Sheikh HF, Tabbara KF. Cataract following posterior chamber phakic intraocular lens. J Refract Surg. 2003; 19(1):72–73. LinkGoogle Scholar
- 11.Sanchez-Galeana CA, Smith RJ, Sanders DR, Lens opacities after posterior chamber phakic intraocular lens implantation. Ophthalmology. 2003; 110(4):781–785.
10.1016/S0161-6420(02)01973-5 Crossref MedlineGoogle Scholar - 12.Gimbel HV, Condon GP, Kohnen T, Olson RJ, Halkiadakis I. Late in-the-bag intraocular lens dislocation: incidence, prevention, and management. J Cataract Refract Surg. 2005; 31(11):2193–2204.
10.1016/j.jcrs.2005.06.053 Crossref MedlineGoogle Scholar - 13.Boxer Wachler BS, Scruggs RT, Yuen LH, Jalali S. Comparison of the Visian ICL and Verisyse phakic intraocular lenses for myopia from 6.00 to 20.00 diopters. J Refract Surg. 2009; 25(9):765–770.
10.3928/1081597X-20090813-02 LinkGoogle Scholar - 14.Kawamorita T, Uozato H, Kamiya K, Shimizu K. Relationship between ciliary sulcus diameter and anterior chamber diameter and corneal diameter. J Cataract Refract Surg. 2010; 36(4):617–624.
10.1016/j.jcrs.2009.11.017 Crossref MedlineGoogle Scholar - 15.Kim KH, Shin HH, Kim HM, Song JS. Correlation between ciliary sulcus diameter measured by 35 MHz ultrasound biomicroscopy and other ocular measurements. J Cataract Refract Surg. 2008; 34(4):632–637.
10.1016/j.jcrs.2007.11.034 Crossref MedlineGoogle Scholar - 16.Fea AM, Annetta F, Cirillo S, Magnetic resonance imaging and Orbscan assessment of the anterior chamber. J Cataract Refract Surg. 2005; 31(9):1713–1718.
10.1016/j.jcrs.2005.02.040 Crossref MedlineGoogle Scholar - 17.Pop M, Payette Y, Mansour M. Predicting sulcus size using ocular measurements. J Cataract Refract Surg. 2001; 27(7):1033–1038.
10.1016/S0886-3350(00)00830-0 Crossref MedlineGoogle Scholar - 18.Werner L, Izak AM, Pandey SK, Apple DJ, Trivedi RH, Schmidbauer JM. Correlation between different measurements within the eye relative to phakic intraocular lens implantation. J Cataract Refract Surg. 2004; 30(9):1982–1988.
10.1016/j.jcrs.2003.10.041 Crossref MedlineGoogle Scholar - 19.Oh J, Shin HH, Kim JH, Kim HM, Song JS. Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy. Ophthalmology. 2007; 114(9):1685–1688.
10.1016/j.ophtha.2006.12.018 Crossref MedlineGoogle Scholar - 20.Biermann J, Bredow L, Boehringer D, Reinhard T. Evaluation of ciliary sulcus diameter using ultrasound biomicroscopy in emmetropic eyes and myopic eyes. J Cataract Refract Surg. 2011; 37(9):1686–1693.
10.1016/j.jcrs.2011.03.048 Crossref MedlineGoogle Scholar - 21.Lee DH, Choi SH, Chung ES, Chung TY. Correlation between preoperative biometry and posterior chamber phakic Visian Implantable Collamer Lens vaulting. Ophthalmology. 2012; 119(2):272–277.
10.1016/j.ophtha.2011.07.047 Crossref MedlineGoogle Scholar - 22.Reinstein DZ, Archer TJ, Silverman RH, Rondeau MJ, Coleman DJ. Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using Artemis VHF digital ultrasound. J Refract Surg. 2009; 25(2):185–194. LinkGoogle Scholar
- 23.Lovisolo CF, Reinstein DZ. Phakic intraocular lenses. Surv Ophthalmol. 2005; 50(6):549–587.
10.1016/j.survophthal.2005.08.011 Crossref MedlineGoogle Scholar - 24.Choi KH, Chung SE, Chung TY, Chung ES. Ultrasound biomicroscopy for determining visian implantable contact lens length in phakic IOL implantation. J Refract Surg. 2007; 23(4):362–367. LinkGoogle Scholar
- 25.Dougherty PJ, Rivera RP, Schneider D, Lane SS, Brown D, Vukich J. Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopy. J Cataract Refract Surg. 2011; 37(1):13–18.
10.1016/j.jcrs.2010.07.014 Crossref MedlineGoogle Scholar - 26.Kojima T, Yokoyama S, Ito M, Horai R, Hara S, Nakamura T, Ichikawa K. Optimization of an Implantable Collamer Lens sizing method using high-frequency ultrasound biomicroscopy. Am J Ophthalmol. 2012; 153(4):632–637.
10.1016/j.ajo.2011.06.031 Crossref MedlineGoogle Scholar - 27.Reinstein DZ, Archer TJ, Silverman RH, Coleman DJ. Accuracy, repeatability, and reproducibility of Artemis very high-frequency digital ultrasound arc-scan lateral dimension measurements. J Cataract Refract Surg. 2006; 32(11):1799–1802.
10.1016/j.jcrs.2006.07.017 Crossref MedlineGoogle Scholar - 28.Pinero DP, Puche AB, Alio JL. Ciliary sulcus diameter and two anterior chamber parameters measured by optical coherence tomography and VHF ultrasound. J Refract Surg. 2009; 25(11):1017–1025.
10.3928/1081597X-20091016-08 LinkGoogle Scholar - 29.Pinero DP, Plaza AB, Alio JL. Anterior segment biometry with 2 imaging technologies: very-high-frequency ultrasound scanning versus optical coherence tomography. J Cataract Refract Surg. 2008; 34(1):95–102.
10.1016/j.jcrs.2007.08.033 Crossref MedlineGoogle Scholar - 30.Sanders DR. Anterior subcapsular opacities and cataracts 5 years after surgery in the visian implantable collamer lens FDA trial. J Refract Surg. 2008; 24(6):566–570. LinkGoogle Scholar