Vector Analysis, Rotational Stability, and Visual Outcome After Implantation of a New Aspheric Toric IOL
Abstract
PURPOSE:
To evaluate vector analysis, rotational stability, and refractive and visual outcome of a new toric intraocular lens (IOL) for correction of preexisting corneal astigmatism during routine cataract surgery.
METHODS:
In this prospective, interventional case series, 30 toric, aspheric Bi-Flex T toric IOLs (Medicontur Medical Engineering Ltd., Inc., Zsámbék, Hungary) were implanted in 20 consecutive patients with topographic corneal astigmatism between 1.50 and 4.00 diopters (D) and evaluated within the first year after implantation. Appropriate IOL-toric alignment was facilitated by combined imaging/eye tracking technology. Postoperative evaluation included refraction and uncorrected and corrected distance visual acuities (UDVA, CDVA). For each visit, photodocumentation in retroillumination was performed to evaluate toric alignment and potential toric IOL rotation. Vector analysis of refractive astigmatism was performed using the Alpins method.
RESULTS:
At 12 months postoperatively, a reduction of the refractive astigmatism from 1.93 ± 0.90 D (range: 0.50 to 4.00 D) to 0.28 ± 0.61 D (range: 0.00 to 1.50 D) could be found, with patients achieving a mean UDVA of 0.06 ± 0.16 logMAR (range: −0.18 to 0.40 logMAR; Snellen 20/20). Intraoperative to 12-month postoperative comparison of IOL axis alignment showed low levels of rotation (0.2° ± 2.41°; range: +4° to −5°). Vector analysis showed target induced astigmatism of 0.60 D @180°, surgically induced astigmatism of 0.80 D @177°, correction index of 1.02 ± 0.25, and a difference vector of 0.30 D @82°.
CONCLUSIONS:
Implantation of the new Bi-Flex T IOL was a safe, stable, and effective method to correct preexisting regular corneal astigmatism during cataract surgery.
[J Refract Surg. 2015;31(8):513–520.]
- 1.Ninn-Pedersen K, Stenevi U, Ehinger B. Cataract patients in a defined Swedish population 1986–1990: II. Preoperative observations. Acta Ophthalmol. 1994; 72:10–15.
10.1111/j.1755-3768.1994.tb02729.x Crossref, Google Scholar - 2.Hoffer KJ. Biometry of 7,500 cataractous eyes. Am J Ophthalmol. 1980; 90:360–368; correction, 890.
10.1016/S0002-9394(14)74917-7 Crossref Medline, Google Scholar - 3.Hoffmann PC, Hütz WW. Analysis of biometry and prevalence data for corneal astigmatism in 23,239 eyes. J Cataract Refract Surg. 2010; 36:1479–1485.
10.1016/j.jcrs.2010.02.025 Crossref Medline, Google Scholar - 4.Guyton DL. Prescribing cylinders: the problem of distortion. Surv Ophthalmol. 1977; 22:177–188.
10.1016/0039-6257(77)90054-6 Crossref Medline, Google Scholar - 5.Mendicute J, Irigoyen C, Ruiz M, Illarramendi I, Ferrer-Blasco T, Montés-Micó R. Toric intraocular lens versus opposite clear corneal incisions to correct astigmatism in eyes having cataract surgery. J Cataract Refract Surg. 2009; 35:451–458.
10.1016/j.jcrs.2008.11.043 Crossref Medline, Google Scholar - 6.De Silva DJ, Ramkissoon YD, Bloom PA. Evaluation of a toric intraocular lens with a Z-haptic. J Cataract Refract Surg. 2006; 32:1492–1498.
10.1016/j.jcrs.2006.04.022 Crossref Medline, Google Scholar - 7.Alberdi T, Macías-Murelaga B, Bascarán L, Goñi N, de Arregui SS, Mendicute J. Rotational stability and visual quality in eyes with Rayner toric intraocular lens implantation. J Refract Surg. 2012; 28:696–701.
10.3928/1081597X-20120921-04 Link, Google Scholar - 8.Sheppard AL, Wolffsohn JS, Bhatt U, Clinical outcomes after implantation of a new hydrophobic acrylic toric IOL during routine cataract surgery. J Cataract Refract Surg. 2013; 39:41–47.
10.1016/j.jcrs.2012.08.055 Crossref Medline, Google Scholar - 9.Entabi M, Harman F, Lee N, Bloom PA. Injectable 1-piece hydrophilic acrylic toric intraocular lens for cataract surgery: efficacy and stability. J Cataract Refract Surg. 2011; 37:235–240.
10.1016/j.jcrs.2010.08.040 Crossref Medline, Google Scholar - 10.Hoffmann PC, Auel S, Hütz WW. Results of higher power toric intraocular lens implantation. J Cataract Refract Surg. 2011; 37:1411–1418.
10.1016/j.jcrs.2011.02.028 Crossref Medline, Google Scholar - 11.Chua WH, Yuen LH, Chua J, Teh G, Hill WE. Matched comparison of rotational stability of 1-piece acrylic and plate-haptic silicone toric intraocular lenses in Asian eyes. J Cataract Refract Surg. 2012; 38:620–624.
10.1016/j.jcrs.2011.10.037 Crossref Medline, Google Scholar - 12.Bachernegg A, Rückl T, Riha W, Grabner G, Dexl AK. Rotational stability and visual outcome after implantation of a new toric intraocular lens for the correction of corneal astigmatism during cataract surgery. J Cataract Refract Surg. 2013; 39:1390–1398.
10.1016/j.jcrs.2013.03.033 Crossref Medline, Google Scholar - 13.Alió JL, Piñero DP, Tomás J, Alesón A. Vector analysis of astigmatic changes after cataract surgery with toric intraocular lens implantation. J Cataract Refract Surg. 2011; 37:1038–1049.
10.1016/j.jcrs.2010.12.053 Crossref Medline, Google Scholar - 14.Alió JL, Agdeppa MC, Pongo VC, El Kady B. Microincision cataract surgery with toric intraocular lens implantation for correcting moderate and high astigmatism: pilot study. J Cataract Refract Surg. 2010; 36:44–52.
10.1016/j.jcrs.2009.07.043 Crossref Medline, Google Scholar - 15.Visser N, Berendschot TT, Bauer NJ, Nuijts RM. Vector analysis of corneal and refractive astigmatism changes following toric pseudophakic and toric phakic IOL implantation. Invest Ophthalmol Vis Sci. 2012; 53:1865–1873.
10.1167/iovs.11-8868 Crossref Medline, Google Scholar - 16.Visser N, Berendschot TT, Bauer NJ, Jurich J, Kersting O, Nuijts RM. Accuracy of toric intraocular lens implantation in cataract and refractive surgery. J Cataract Refract Surg. 2011; 37:1394–1402.
10.1016/j.jcrs.2011.02.024 Crossref Medline, Google Scholar - 17.Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993; 19:524–533.
10.1016/S0886-3350(13)80617-7 Crossref Medline, Google Scholar - 18.Alpins N. Astigmatism analysis by the Alpins method. J Cataract Refract Surg. 2001; 27:31–49.
10.1016/S0886-3350(00)00798-7 Crossref Medline, Google Scholar - 19.Weinand F, Jung A, Stein A, Pfotzner A, Becker R, Pavlovic S. Rotational stability of a single-piece hydrophobic acrylic intraocular lens: new method for high-precision rotation control. J Cataract Refract Surg. 2007; 33:800–803.
10.1016/j.jcrs.2007.01.030 Crossref Medline, Google Scholar - 20.Becker KA, Auffarth GU, Völcker HE. Measurement method for the determination of rotation and decentration of intraocular lenses [article in German]. Ophthalmologe. 2004; 101:600–603. Medline, Google Scholar
- 21.Carvalho MJ, Suzuki SH, Freitas LL, Branco BC, Schor P, Lima AL. Limbal relaxing incisions to correct corneal astigmatism during phacoemulsification. J Refract Surg. 2007; 23:499–504. Link, Google Scholar
- 22.Horn JD. Status of toric intraocular lenses. Curr Opin Ophthalmol. 2007; 18:58–61.
10.1097/ICU.0b013e328011f9bf Crossref Medline, Google Scholar - 23.Goggin M, Moore S, Esterman A. Toric intraocular lens outcome using the manufacturer’s prediction of corneal plane equivalent intraocular lens cylinder power. Arch Ophthalmol. 2011; 129:1004–1008.
10.1001/archophthalmol.2011.178 Crossref Medline, Google Scholar