Clinical Evaluation of Reading Performance Using the Salzburg Reading Desk With a Refractive Rotational Asymmetric Multifocal Intraocular Lens
Abstract
PURPOSE:
To evaluate functional results and reading performance using the Salzburg Reading Desk after implantation of a sector-shaped near-embedded, rotational asymmetrical multifocal intraocular lens (IOL) and a multifocal toric IOL with a +3.00 diopter (D) near addition.
METHODS:
In a prospective study, the LentisMplus and Mplus toric IOLs (Oculentis GmbH, Berlin, Germany) were implanted in 34 eyes of 18 patients at the University Eye Hospital of Heidelberg. Uncorrected and corrected distance visual acuity (UDVA, CDVA) and uncorrected and corrected near visual acuity (UNVA, CNVA) were evaluated using standardized visual acuity charts (ETDRS). The Salzburg Reading Desk was used to analyze unilateral and bilateral uncorrected and corrected reading acuity, reading distance, reading speed, and the smallest log-scaled print size that could be read effectively at a set (40 cm/80 cm) and subjective chosen near and intermediate distance.
RESULTS:
Postoperatively, the median UDVA was 0.08 logMAR (20/25 Snellen) and the median CDVA was 0.01 logMAR (20/20 Snellen). The median UNVA was 0.12 logMAR (20/25 Snellen) and the median CNVA was 0.03 logMAR (20/20 Snellen). The median uncorrected reading acuity measured with the Salzburg Reading Desk for near distance at 40 cm was 0.18 logMAR (20/32 Snellen). The subjectively preferred near distance was 39 cm and revealed similar visual acuity results. The best reading acuity for intermediate distance with a median of 0.22 logMAR (20/32 Snellen) was achieved at a median distance of 62 cm.
CONCLUSIONS:
Reading performance of the multifocal IOL corresponded for near standardized and individual distance, whereas reading function was better at the patient's preferred intermediate distance.
[J Refract Surg. 2016;32(8):526–532.]
- 1.Auffarth GU, Apple DJ. History of the development of intraocular lenses [article in German]. Ophthalmologe. 2001; 98:1017–1028.
10.1007/s003470170020 Crossref Medline, Google Scholar - 2.Auffarth GU, Dick HB. Multifocal intraocular lenses: a review [article in German]. Ophthalmologe. 2001; 98:127–137.
10.1007/s003470170172 Crossref Medline, Google Scholar - 3.Khoramnia R, Auffarth GU, Rabsilber TM, Holzer MP. Implantation of a multifocal toric intraocular lens with a surface-embedded near segment after repeated LASIK treatments. J Cataract Refract Surg. 2012; 38:2049–2052.
10.1016/j.jcrs.2012.08.042 Crossref Medline, Google Scholar - 4.Kohnen T, Nuijts R, Levy P, Haefliger E, Alfonso JF. Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a +3.0 D addition. J Cataract Refract Surg. 2009; 35:2062–2069.
10.1016/j.jcrs.2009.08.013 Crossref Medline, Google Scholar - 5.de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: literature review of benefits and side effects. J Cataract Refract Surg. 2013; 39:268–278.
10.1016/j.jcrs.2012.12.002 Crossref Medline, Google Scholar - 6.Venter JA, Pelouskova M, Bull CE, Schallhorn SC, Hannan SJ. Visual outcomes and patient satisfaction with a rotational asymmetric refractive intraocular lens for emmetropic presbyopia. J Cataract Refract Surg. 2015; 41:585–593.
10.1016/j.jcrs.2014.06.035 Crossref Medline, Google Scholar - 7.Venter JA, Pelouskova M, Collins BM, Schallhorn SC, Hannan SJ. Visual outcomes and patient satisfaction in 9366 eyes using a refractive segmented multifocal intraocular lens. J Cataract Refract Surg. 2013; 39:1477–1484.
10.1016/j.jcrs.2013.03.035 Crossref Medline, Google Scholar - 8.Thomas BC, Auffarth GU, Philips R, Clinical results after implantation of a new segmental refractive multifocal intraocular lens [article in German]. Ophthalmologe. 2013; 110:1058–1062.
10.1007/s00347-013-2810-5 Crossref Medline, Google Scholar - 9.McAlinden C, Moore JE. Multifocal intraocular lens with a surface-embedded near section: short-term clinical outcomes. J Cataract Refract Surg. 2011; 37:441–445.
10.1016/j.jcrs.2010.08.055 Crossref Medline, Google Scholar - 10.Alió JL, Piñero DP, Plaza-Puche AB, Chan MJ. Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lens. J Cataract Refract Surg. 2011; 37:241–250.
10.1016/j.jcrs.2010.08.043 Crossref Medline, Google Scholar - 11.Dexl AK, Schlögel H, Wolfbauer M, Grabner G. Device for improving quantification of reading acuity and reading speed. J Refract Surg. 2010; 26:682–688.
10.3928/1081597X-20091119-01 Link, Google Scholar - 12.Alió JL, Plaza-Puche AB, Piñero DP, Javaloy J, Ayala MJ. Comparative analysis of the clinical outcomes with 2 multifocal intraocular lens models with rotational asymmetry. J Cataract Refract Surg. 2011; 37:1605–1614.
10.1016/j.jcrs.2011.03.054 Crossref Medline, Google Scholar - 13.Muñoz G, Albarrán-Diego C, Javaloy J, Sakla HF, Cerviño A. Combining zonal refractive and diffractive aspheric multifocal intraocular lenses. J Refract Surg. 2012; 28:174–181.
10.3928/1081597X-20120215-02 Link, Google Scholar - 14.Muñoz G, Albarrán-Diego C, Ferrer-Blasco T, Sakla HF, García Lázaro S. Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens. J Cataract Refract Surg. 2011; 37:2043–2052.
10.1016/j.jcrs.2011.05.045 Crossref Medline, Google Scholar - 15.Alió JL, Plaza-Puche AB, Javaloy J, Ayala MJ, Moreno LJ, Piñero DP. Comparison of a new refractive multifocal intraocular lens with an inferior segmental near add and a diffractive multifocal intraocular lens. Ophthalmology. 2012; 119:555–563.
10.1016/j.ophtha.2011.08.036 Crossref Medline, Google Scholar - 16.Alió JL, Plaza-Puche AB, Javaloy J, Ayala MJ. Comparison of the visual and intraocular optical performance of a refractive multifocal IOL with rotational asymmetry and an apodized diffractive multifocal IOL. J Refract Surg. 2012; 28:100–105.
10.3928/1081597X-20120110-01 Link, Google Scholar - 17.Alió JL, Plaza-Puche AB, Javaloy J, Ayala MJ, Vega-Estrada A. Clinical and optical intraocular performance of rotationally asymmetric multifocal IOL plate-haptic design versus C-loop haptic design. J Refract Surg. 2013; 29:252–259.
10.3928/1081597X-20130318-04 Link, Google Scholar - 18.De Vries NE, Webers CA, Montés-Micó R, Long-term follow-up of a multifocal apodized diffractive intraocular lens after cataract surgery. J Cataract Refract Surg. 2008; 34:1476–1482.
10.1016/j.jcrs.2008.05.030 Crossref Medline, Google Scholar - 19.Kohnen T, Nuijts R, Levy P, Haefliger E, Alfonso JF. Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a +3.0 D addition. J Cataract Refract Surg. 2009; 35:2062–2069.
10.1016/j.jcrs.2009.08.013 Crossref Medline, Google Scholar - 20.Cillino S, Casuccio A, Di Pace F, One-year outcomes with new-generation multifocal intraocular lenses. Ophthalmology. 2008; 115:1508–1516.
10.1016/j.ophtha.2008.04.017 Crossref Medline, Google Scholar - 21.Radner W, Obermayer W, Richter-Mueksch S, Willinger U, Velikay-Parel M, Eisenwort B. The validity and reliability of short German sentences for measuring reading speed. Graefes Arch Clin Exp Ophthalmol. 2002; 240:461–467.
10.1007/s00417-002-0443-5 Crossref Medline, Google Scholar - 22.Dexl A, Schlögel H, Wolfbauer M, Die Entwicklung einer neuen Methode zur Bestimmung der Leseschärfe – Das “Salzburg Reading Desk (SRD).”Spektrum Augenheilkd. 2009; 23:435–438.
10.1007/s00717-009-0370-5 Crossref, Google Scholar - 23.Alió JL, Grabner G, Plaza-Puche AB, Postoperative bilateral reading performance with 4 intraocular lens models: six-months results. J Cataract Refract Surg. 2011; 37:842–852.
10.1016/j.jcrs.2010.11.039 Crossref Medline, Google Scholar - 24.Rasp M, Bachernegg A, Seyeddain O, Bilateral reading performance of 4 multifocal intraocular lens models and a monofocal intraocular lens under bright lighting conditions. J Cataract Refract Surg. 2012; 38:1950–1961.
10.1016/j.jcrs.2012.07.027 Crossref Medline, Google Scholar - 25.Santhiago MR, Netto MV, Espindola RF, Comparison of reading performance after bilateral implantation of multifocal intraocular lenses with +3.00 or +4.00 diopter addition. J Cataract Refract Surg. 2010; 36:1874–1879.
10.1016/j.jcrs.2010.05.022 Crossref Medline, Google Scholar