Toric ICL Implantation After Sequential Intracorneal Ring Segments Implantation and Corneal Cross-linking in Keratoconus: 2-Year Follow-up
Abstract
PURPOSE:
To evaluate 6-month and 2-year safety and clinical outcomes of Visian toric Implantable Collamer Lens (toric ICL) (STAAR Surgical, Monrovia, CA) implantation for the treatment of residual refractive errors after sequential intracorneal ring segments (ICRS) insertion and cross-linking (CXL) in keratoconus.
METHODS:
This consecutive case series included the results of a three-step ICRS–CXL–toric ICL procedure in 16 eyes of 13 patients with moderate to severe keratoconus (stages II and III of Amsler–Krumeich classification). The ICRS and CXL procedures were performed sequentially with an interval of 4 weeks and the toric ICL implantation was performed at least 6 months after CXL. Data were collected preoperatively, at 6 months after sequential ICRS–CXL, and at 6 and 24 months following toric ICL implantation. All 16 eyes were evaluated at the 2-year follow-up.
RESULTS:
There was a significant decrease in keratometry (steep, flat, and maximum) and refraction (sphere and spherical equivalence, but not cylinder) and a significant increase in both uncorrected (UDVA) (from 1.06 to 0.76 logMAR, P = .004) and corrected (CDVA) (from 0.42 to 0.26 logMAR; P = .002) distance visual acuity 6 months following sequential ICRS–CXL, whereas UDVA and refraction significantly improved 6 months after ICL insertion (UDVA reached 0.33 logMAR, P = .001). At the 24-month follow-up, UDVA and keratometric readings were stable, whereas CDVA, sphere, and cylinder showed a significant improvement.
CONCLUSIONS:
Implantation of Visian toric ICL following sequential ICRS insertion and CXL is an effective and safe option for correcting high residual refractive error and improving visual acuity in patients with moderate to severe keratoconus in the long term.
[J Refract Surg. 2017;33(9):610–616.]
- 1.Krachmer JH, Feder RS, Belin MW. Keratoconus and related noninflammatory corneal thinning disorders. Surv Ophthalmol. 1984; 28:293–322.
10.1016/0039-6257(84)90094-8 Crossref Medline, Google Scholar - 2.Rabinowitz YS. Keratoconus. Surv Ophthalmol. 1998; 42:297–319.
10.1016/S0039-6257(97)00119-7 Crossref Medline, Google Scholar - 3.Wollensak G, Spoerl E, Seiler T. Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus. Am J Ophthalmol. 2003; 135:620–627.
10.1016/S0002-9394(02)02220-1 Crossref Medline, Google Scholar - 4.Kamburoglu G, Ertan A. Intacs implantation with sequential collagen cross-linking treatment in postoperative LASIK ectasia. J Refract Surg. 2008; 24:S726–S729. Medline, Google Scholar
- 5.Chan CCK, Sharma M, Wachler BSB. Effect of inferior-segment Intacs with and without C3-R on keratoconus. J Cataract Refract Surg. 2007; 33:75–80.
10.1016/j.jcrs.2006.09.012 Crossref Medline, Google Scholar - 6.Kremer I, Aizenman I, Lichter H, Shayer S, Levinger S. Simultaneous wavefront-guided photorefractive keratectomy and corneal collagen crosslinking after intrastromal corneal ring segment implantation for keratoconus. J Cataract Refract Surg. 2012; 38:1802–1807.
10.1016/j.jcrs.2012.05.033 Crossref Medline, Google Scholar - 7.Fadlallah A, Dirani A, Chelala E, Antonios R, Cherfan G, Jarade E. Non-topography-guided PRK combined with CXL for the correction of refractive errors in patients with early stage keratoconus. J Refract Surg. 2014; 30:688.
10.3928/1081597X-20140903-02 Link, Google Scholar - 8.Kymionis GD, Grentzelos MA, Karavitaki AE, Zotta P, Yoo SH, Pallikaris IG. Combined corneal collagen cross-linking and posterior chamber toric implantable collamer lens implantation for keratoconus. Ophthalmic Surg Lasers Imaging, 2011; 42:e22–e25. Link, Google Scholar
- 9.Fadlallah A, Dirani A, El Rami H, Cherfane G, Jarade E. Safety and visual outcome of Visian toric ICL implantation after corneal collagen cross-linking in keratoconus. J Refract Surg. 2013; 29:84–89.
10.3928/1081597X-20130117-01 Link, Google Scholar - 10.Antonios R, Dirani A, Fadlallah A, Safety and visual outcome of Visian toric ICL implantation after corneal collagen cross-linking in keratoconus: up to 2 years of follow-up. J Ophthalmol. 2015; 2015:514834.
10.1155/2015/514834 Crossref Medline, Google Scholar - 11.Fadlallah A, Dirani A, El Rami H, Cherfane G, Jarade E. Safety and visual outcome of Visian toric ICL implantation after corneal collagen cross-linking in keratoconus. J Refract Surg. 2013; 29:84–89.
10.3928/1081597X-20130117-01 Link, Google Scholar - 12.Dirani A, Fadlallah A, Khoueir Z, Antoun J, Cherfan G, Jarade E. Visian toric ICL implantation after intracorneal ring segments implantation and corneal collagen crosslinking in keratoconus. Eur J Ophthalmol. 2013; 24:338–344.
10.5301/ejo.5000384 Crossref Medline, Google Scholar - 13.Le Loir M, Cochener B. Long-term results of posterior chamber phakic intraocular lens implantation for correction of high ametropia [article in French]. J Fr Ophtalmol. 2012; 35:402–411.
10.1016/j.jfo.2011.06.006 Crossref Medline, Google Scholar - 14.Kamiya K, Shimizu K, Igarashi A, Hikita F, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009; 127:845–850.
10.1001/archophthalmol.2009.67 Crossref Medline, Google Scholar - 15.Kamiya K, Shimizu K, Kobashi H, Komatsu M, Nakamura A, Nakamura T. Clinical outcomes of posterior chamber toric phakic intraocular lens implantation for the correction of high myopic astigmatism in eyes with keratoconus: 6-month follow-up. Graefes Arch Clin Exp Ophthalmol. 2011; 249:1073–1080.
10.1007/s00417-010-1540-5 Crossref Medline, Google Scholar - 16.Izquierdo L, Henriquez MA, McCarthy M. Artiflex phakic intraocular lens implantation after corneal collagen cross-linking in keratoconic eyes. J Refract Surg. 2011; 27:482–487.
10.3928/1081597X-20101223-02 Link, Google Scholar - 17.Alfonso JF, Fernández-Vega L, Lisa C, Fernandes P, González-Méijome JM, Montés-Micó R. Collagen copolymer toric posterior chamber phakic intraocular lens in eyes with keratoconus. J Cataract Refract Surg. 2010; 36:906–916.
10.1016/j.jcrs.2009.11.032 Crossref Medline, Google Scholar - 18.Coskunseven E, Onder M, Kymionis GD, Diakonis VF, Arslan E, Tsiklis N. Combined Intacs and posterior chamber toric implantable Collamer lens implantation for keratoconic patients with extreme myopia. Am J Ophthalmol. 2007; 144:387–389.
10.1016/j.ajo.2007.05.035 Crossref Medline, Google Scholar - 19.Alfonso JF, Lisa C, Fernández-Vega L, Madrid-Costa D, Poo-López A, Montés-Micó R. Intrastromal corneal ring segments and posterior chamber phakic intraocular lens implantation for keratoconus correction. J Cataract Refract Surg. 2011; 37:706–713.
10.1016/j.jcrs.2010.10.060 Crossref Medline, Google Scholar - 20.El-Raggal TM, Abdel Fattah AA. Sequential Intacs and Verisyse phakic intraocular lens for refractive improvement in keratoconic eyes. J Cataract Refract Surg. 2007; 33:966–970.
10.1016/j.jcrs.2007.02.024 Crossref Medline, Google Scholar - 21.Coskunseven E, Sharma DP, Jankov MR, Kymionis GD, Richoz O, Hafezi F. Collagen copolymer toric phakic intraocular lens for residual myopic astigmatism after intrastromal corneal ring segment implantation and corneal collagen crosslinking in a 3-stage procedure for keratoconus. J Cataract Refract Surg. 2013; 39:722–729.
10.1016/j.jcrs.2012.11.027 Crossref Medline, Google Scholar - 22.Navas A, Tapia-Herrera G, Jaimes M, Implantable collamer lenses after intracorneal ring segments for keratoconus. Int Ophthalmol. 2012; 32:423–429.
10.1007/s10792-012-9582-9 Crossref Medline, Google Scholar - 23.Grewal DS, Brar GS, Jain R, Sood V, Singla M, Grewal SPS. Corneal collagen crosslinking using riboflavin and ultraviolet- A light for keratoconus: one-year analysis using Scheimpflug imaging. J Cataract Refract Surg. 2009; 35:425–432.
10.1016/j.jcrs.2008.11.046 Crossref Medline, Google Scholar - 24.Caporossi A, Mazzotta C, Baiocchi S, Caporossi T. Long-term results of riboflavin ultraviolet a corneal collagen cross-linking for keratoconus in Italy: the Siena eye cross study. Am J Ophthalmol. 2010; 149:585–593.
10.1016/j.ajo.2009.10.021 Crossref Medline, Google Scholar - 25.Hashem AN, El Danasoury AM, Anwar HM. Axis alignment and rotational stability after implantation of the toric implantable collamer lens for myopic astigmatism. J Refract Surg. 2009; 25:S939–S9243.
10.3928/1081597X-20090915-08 Link, Google Scholar