Abstract

PURPOSE:

To report the outcomes of laser in situ keratomileusis (LASIK) for hyperopia using the Triple-A ablation profile with the MEL 90 excimer laser (Carl Zeiss Meditec, Jena, Germany).

METHODS:

This retrospective analysis included 1,383 eyes treated by LASIK for hyperopia using the Triple-A ablation profile with the MEL 90 at London Vision Clinic, London, United Kingdom, between September 2013 and December 2016. Inclusion criteria were attempted hyperopic correction of +0.25 diopters (D) or higher and corrected distance visual acuity (CDVA) of 20/40 or better. Patients were observed for 1 year after surgery. Standard outcomes analysis was performed.

RESULTS:

One-year data were available for 1,350 (97%) eyes. Mean attempted spherical equivalent refraction (SEQ) was +2.77 ± 1.34 D (range: +0.13 to +6.50 D) and mean cylinder was −0.67 ± 0.66 D (range: 0.00 to −5.00 D). Mean age was 54 ± 11 years (range: 21 to 75 years), and 57% were female. Postoperative spherical equivalent was ±0.50 D in 73% and ±1.00 D in 93% of eyes. Uncorrected distance visual acuity was 20/20 or better in 75% of eyes, relative to 93% with preoperative CDVA of 20/20 or better. One line of CDVA was lost in 17% of eyes and two lines were lost in 0.6% of eyes. There was a clinically insignificant but statistically significant increase (P < .01) in contrast sensitivity at 3 and 6 cycles per degree (cpd) and no change for 12 and 18 cpd.

CONCLUSIONS:

LASIK for hyperopia with the MEL 90 excimer laser was found to satisfy accepted criteria for safety, efficacy, and stability.

[J Refract Surg. 2018;34(12):799–808.]

  • 1.Barraquer JI. Queratomileusis y queratofakia. Bogota: Instituto Barraquer de America, 1980:342.

    Google Scholar
  • 2.Chayet AS, Assil KK, Montes M, Castellanos A. Laser in situ keratomileusis for hyperopia: new software. J Refract Surg. 1997; 13:S434–S435.

    LinkGoogle Scholar
  • 3.Ojeimi G, Waked N. Laser in situ keratomileusis for hyperopia. J Refract Surg. 1997; 13:S432–S433.

    LinkGoogle Scholar
  • 4.Dausch D, Smecka Z, Klein R, Schroder E, Kirchner S. Excimer laser photorefractive keratectomy for hyperopia. J Cataract Refract Surg. 1997; 23:169–176.10.1016/S0886-3350(97)80338-0

    Crossref MedlineGoogle Scholar
  • 5.Arbelaez MC, Knorz MC. Laser in situ keratomileusis for hyperopia and hyperopic astigmatism. J Refract Surg. 1999; 15:406–414.

    LinkGoogle Scholar
  • 6.Esquenazi S, Mendoza A. Two-year follow-up of laser in situ keratomileusis for hyperopia. J Refract Surg. 1999; 15:648–652.

    LinkGoogle Scholar
  • 7.Argento CJ, Cosentino MJ. Comparison of optical zones in hyperopic laser in situ keratomileusis: 5.9 mm versus smaller optical zones. J Cataract Refract Surg. 2000; 26:1137–1146.10.1016/S0886-3350(99)00356-9

    Crossref MedlineGoogle Scholar
  • 8.Cobo-Soriano R, Llovet F, Gonzalez-Lopez F, Domingo B, Gomez-Sanz F, Baviera J. Factors that influence outcomes of hyperopic laser in situ keratomileusis. J Cataract Refract Surg. 2002; 28:1530–1538.10.1016/S0886-3350(02)01367-6

    Crossref MedlineGoogle Scholar
  • 9.Reinstein DZ, Couch DG, Archer TJ. LASIK for hyperopic astigmatism and presbyopia using micro-monovision with the Carl Zeiss Meditec MEL80. J Refract Surg. 2009; 25:37–58.

    LinkGoogle Scholar
  • 10.Reinstein DZ, Carp GI, Archer TJ, et al.LASIK for the correction of high hyperopic astigmatism with epithelial thickness monitoring. J Refract Surg. 2017; 33:314–321.10.3928/1081597X-20170111-04

    LinkGoogle Scholar
  • 11.Waring GO, Fant B, Stevens G, et al.Laser in situ keratomileusis for spherical hyperopia and hyperopic astigmatism using the NIDEK EC-5000 excimer laser. J Refract Surg. 2008; 24:123–136.

    LinkGoogle Scholar
  • 12.Kermani O, Oberheide U, Schmiedt K, Gerten G, Bains HS. Outcomes of hyperopic LASIK with the NIDEK NAVEX platform centered on the visual axis or line of sight. J Refract Surg. 2009; 25:S98–S103.

    LinkGoogle Scholar
  • 13.Alió J, Galal A, Ayala MJ, Artola A. Hyperopic LASIK with Esiris/Schwind technology. J Refract Surg. 2006; 22:772–781.

    LinkGoogle Scholar
  • 14.de Ortueta D, Arba-Mosquera S, Baatz H. Topographic changes after hyperopic LASIK with the SCHWIND ESIRIS laser platform. J Refract Surg. 2008; 24:137–144.

    LinkGoogle Scholar
  • 15.Arba-Mosquera S, de Ortueta D. LASIK for hyperopia using an aberration-neutral profile with an asymmetric offset centration. J Refract Surg. 2016; 32:78–83.10.3928/1081597X-20151119-04

    LinkGoogle Scholar
  • 16.Kanellopoulos AJ, Conway J, Pe LH. LASIK for hyperopia with the WaveLight excimer laser. J Refract Surg. 2006; 22:43–47.

    LinkGoogle Scholar
  • 17.Kanellopoulos AJ. Topography-guided hyperopic and hyper-opic astigmatism femtosecond laser-assisted LASIK: long-term experience with the 400 Hz eye-Q excimer platform. Clin Ophthalmol. 2012; 6:895–901.10.2147/OPTH.S23573

    Crossref MedlineGoogle Scholar
  • 18.Leccisotti A. Femtosecond laser-assisted hyperopic laser in situ keratomileusis with tissue-saving ablation: analysis of 800 eyes. J Cataract Refract Surg. 2014; 40:1122–1130.10.1016/j.jcrs.2013.11.031

    Crossref MedlineGoogle Scholar
  • 19.de Ortueta D, Arba-Mosquera S. Laser in situ keratomileusis for high hyperopia with corneal vertex centration and asymmetric offset. Eur J Ophthalmol. 2016; 27:141–152.10.5301/ejo.5000835

    Crossref MedlineGoogle Scholar
  • 20.Plaza-Puche AB, Aswad AE, Arba-Mosquera S, Wrobel-Dudzinska D, Abdou AA, Alió JL. Optical profile following high hyperopia correction with a 500-Hz excimer laser system. J Refract Surg. 2016; 32:6–13.10.3928/1081597X-20151207-06

    LinkGoogle Scholar
  • 21.Amigó A, Bonaque-González S, Guerras-Valera E. Control of induced spherical aberration in moderate hyperopic LASIK by customizing corneal asphericity. J Refract Surg. 2015; 31:802–806.10.3928/1081597X-20151111-03

    LinkGoogle Scholar
  • 22.Plaza-Puche AB, Yebana P, Arba-Mosquera S, Alió JL. Three-year follow-up of hyperopic LASIK using a 500-Hz excimer laser system. J Refract Surg. 2015; 31:674–682.10.3928/1081597X-20150928-06

    LinkGoogle Scholar
  • 23.Antonios R, Arba Mosquera S, Awwad ST. Hyperopic laser in situ keratomileusis: comparison of femtosecond laser and mechanical microkeratome flap creation. J Cataract Refract Surg. 2015; 41:1602–1609.10.1016/j.jcrs.2014.11.049

    Crossref MedlineGoogle Scholar
  • 25.Alió JL, El Aswad A, Vega-Estrada A, Javaloy J. Laser in situ keratomileusis for high hyperopia (>5.0 diopters) using optimized aspheric profiles: efficacy and safety. J Cataract Refract Surg. 2013; 39:519–527.10.1016/j.jcrs.2012.10.045

    Crossref MedlineGoogle Scholar
  • 25.U.S. Food and Drug Administration. Carl Zeiss Meditec Inc. MEL 80 Excimer Laser System (Hyperopia) (PMA). http://www.accessdata.fda.gov/cdrh_docs/pdf6/P060004b.pdf. Accessed October 19, 2018.

    Google Scholar
  • 26.El-Naggar MT, Hovaghimian DG. Assessment of refractive outcome of femtosecond-assisted LASIK for hyperopia correction. Electron Physician. 2017; 9:3958–3965.10.19082/3958

    Crossref MedlineGoogle Scholar
  • 27.Garcia-Gonzalez M, Iglesias-Iglesias M, Drake Rodriguez-Casanova P, Gros-Otero J, Teus MA. Femtosecond laser-assisted LASIK with and without the adjuvant use of mitomycin C to correct hyperopia. J Refract Surg. 2018; 34:23–28.10.3928/1081597X-20171116-01

    LinkGoogle Scholar
  • 28.Kermani O, Schmeidt K, Oberheide U, Gerten G. Hyperopic laser in situ keratomileusis with 5.5-, 6.5-, and 7.0-mm optical zones. J Refract Surg. 2005; 21:52–58.

    LinkGoogle Scholar
  • 29.Reinstein DZ, Carp GI, Lewis TA, Archer TJ, Gobbe M. Outcomes for myopic LASIK with the MEL 90 excimer laser. J Refract Surg. 2015; 31:316–321.10.3928/1081597X-20150423-05

    LinkGoogle Scholar
  • 30.Reinstein DZ, Yap TE, Carp GI, Archer TJ, Gobbe M. Reproducibility of manifest refraction between surgeons and optometrists in a clinical refractive surgery practice. J Cataract Refract Surg. 2014; 40:450–459.10.1016/j.jcrs.2013.08.053

    Crossref MedlineGoogle Scholar
  • 31.Mastropasqua L, Calienno R, Lanzini M, et al.Opaque bubble layer incidence in femtosecond laser-assisted LASIK: comparison among different flap design parameters. Int Ophthalmol. 2017; 37:635–641.10.1007/s10792-016-0323-3

    Crossref MedlineGoogle Scholar
  • 32.Wu N, Christenbury JG, Dishler JG, et al.A technique to reduce incidence of opaque bubble layer formation during LASIK flap creation using the VisuMax femtosecond laser. J Refract Surg. 2017; 33:584–590.10.3928/1081597X-20170621-06

    LinkGoogle Scholar
  • 33.Reinstein DZ, Gobbe M, Archer TJ. Coaxially sighted corneal light reflex versus entrance pupil center centration of moderate to high hyperopic corneal ablations in eyes with small and large angle kappa. J Refract Surg. 2013; 29:518–525.10.3928/1081597X-20130719-08

    LinkGoogle Scholar
  • 34.Chang DH, Waring GO. The subject-fixated coaxially sighted corneal light reflex: a clinical marker for centration of refractive treatments and devices. Am J Ophthalmol. 2014; 158:863–874.10.1016/j.ajo.2014.06.028

    Crossref MedlineGoogle Scholar
  • 35.Moshirfar M, Hoggan RN, Muthappan V. Angle kappa and its importance in refractive surgery. Oman J Ophthalmol. 2013; 6:151–158.10.4103/0974-620X.122268

    Crossref MedlineGoogle Scholar
  • 36.Park CY, Oh SY, Chuck RS. Measurement of angle kappa and centration in refractive surgery. Curr Opin Ophthalmol. 2012; 23:269–275.10.1097/ICU.0b013e3283543c41

    Crossref MedlineGoogle Scholar
  • 37.Arba Mosquera S, Verma S, McAlinden C. Centration axis in refractive surgery. Eye Vis (Lond). 2015; 2:4.10.1186/s40662-015-0014-6

    Crossref MedlineGoogle Scholar
  • 38.Reinstein DZ, Carp GI, de Benedictis D, et al.Standardization of LASIK surgical technique evaluated by comparison of procedure time between two experienced surgeons. J Cataract Refract Surg. 2015; 41:1004–1008.10.1016/j.jcrs.2014.08.039

    Crossref MedlineGoogle Scholar
  • 39.Reinstein DZ, Archer TJ, Randleman JB. JRS Standard for reporting astigmatism outcomes of refractive surgery. J Refract Surg. 2014; 30:654–659.10.3928/1081597X-20140903-01

    LinkGoogle Scholar
  • 40.Alpins N. Astigmatism analysis by the Alpins method. J Cataract Refract Surg. 2001; 27:31–49.10.1016/S0886-3350(00)00798-7

    Crossref MedlineGoogle Scholar
  • 41.Reinstein DZ, Archer TJ, Gobbe M, Silverman RH, Coleman DJ. Epithelial thickness after hyperopic LASIK: three-dimensional display with Artemis very high-frequency digital ultrasound. J Refract Surg. 2010; 26:555–564.10.3928/1081597X-20091105-02

    LinkGoogle Scholar
  • 42.Arba-Mosquera S, de Ortueta D. Geometrical analysis of the loss of ablation efficiency at non-normal incidence. Opt Express. 2008; 16:3877–3895.10.1364/OE.16.003877

    Crossref MedlineGoogle Scholar
  • 43.Mrochen M, Seiler T. Influence of corneal curvature on calculation of ablation patterns used in photorefractive laser surgery. J Refract Surg. 2001; 17:S584–S587.

    LinkGoogle Scholar
  • 44.Reinstein DZ, Archer TJ, Gobbe M, Silverman R, Coleman DJ. Stromal thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound. J Refract Surg. 2009; 25:776–786.10.3928/1081597X-20090813-04

    LinkGoogle Scholar
  • 45.Guzowski M, Wang JJ, Rochtchina E, Rose KA, Mitchell P. Five-year refractive changes in an older population: the Blue Mountains Eye Study. Ophthalmology. 2003; 110:1364–1370.10.1016/S0161-6420(03)00465-2

    Crossref MedlineGoogle Scholar

We use cookies on this site to enhance your user experience. For a complete overview of all the cookies used, please see our privacy policy.

×